Monitoring Of Stérilisateur ETO
Effective monitoring of the Oxyde d'éthylène (S'ALIGNER) stérilisation process is essential to guarantee the safety of medical devices. Un système de contrôle qualité robuste intègre des, chimique, et des indicateurs biologiques pour garantir que chaque lot répond aux niveaux internationaux d'assurance de stérilité.
- Continuous recording of Temperature (37-55℃).
- Pressure and Humidity (40%-60%RH) surveillance.
- Exposure time (1-3h) and aeration/exhaust time (up to 12h) tracking.
- Compliance with equipment operation manuals.
- External indicators on every package to confirm process exposure.
- Internal indicator cards placed in the “most-difficult-to-sterilize” locations.
- Verification of color change (typically from Brown to Green).
- Performed for every sterilization batch.
- Utilizes Bacillus stearothermophilus spore variants.
- Indicators are cultured immediately post-cycle.
- The gold standard for determining absolute sterilization efficacy.
4. Post-sterilization Treatment
Once the cycle is complete, strict handling protocols must be followed to maintain sterility: