Effective monitoring of the Etilen Oksida (BERBARIS) sterilisasi process is essential to guarantee the safety of medical devices. Sistem kendali mutu yang kuat mengintegrasikan fisik, kimia, dan indikator biologis untuk memastikan setiap batch memenuhi tingkat jaminan sterilitas internasional.
- Continuous recording of Temperature (37-55℃).
- Pressure and Humidity (40%-60%RH) pemantauan.
- Exposure time (1-3H) and aeration/exhaust time (up to 12h) tracking.
- Compliance with equipment operation manuals.
- External indicators on every package to confirm process exposure.
- Internal indicator cards placed in the “most-difficult-to-sterilize” locations.
- Verification of color change (typically from Brown to Green).
- Performed for every sterilization batch.
- Utilizes Bacillus stearothermophilus spore variants.
- Indicators are cultured immediately post-cycle.
- The gold standard for determining absolute sterilization efficacy.
4. Post-sterilization Treatment
Once the cycle is complete, strict handling protocols must be followed to maintain sterility: