Monitoring Of ETO-sterilisator
Effective monitoring of the Etileenoksied (LINE UP) sterilisasie process is essential to guarantee the safety of medical devices. 'N robuuste kwaliteit beheer stelsel integreer fisiese, chemiese, en biologiese aanwysers om te verseker dat elke bondel aan internasionale steriliteitsversekeringsvlakke voldoen.
- Continuous recording of Temperature (37-55℃).
- Pressure and Humidity (40%-60%RH) monitoring.
- Exposure time (1-3h) and aeration/exhaust time (up to 12h) tracking.
- Compliance with equipment operation manuals.
- External indicators on every package to confirm process exposure.
- Internal indicator cards placed in the “most-difficult-to-sterilize” locations.
- Verification of color change (typically from Brown to Green).
- Performed for every sterilization batch.
- Utilizes Bacillus stearothermophilus spore variants.
- Indicators are cultured immediately post-cycle.
- The gold standard for determining absolute sterilization efficacy.
4. Post-sterilization Treatment
Once the cycle is complete, strict handling protocols must be followed to maintain sterility: