Monitoring Of ETO Sterilizer
Effective monitoring of the Ethylene oxide (Pumila) isterilisasyon process is essential to guarantee the safety of medical devices. Ang isang matatag na sistema ng kontrol sa kalidad ay nagsasama ng pisikal, kemikal, at mga biological indicator upang matiyak na ang bawat batch ay nakakatugon sa internasyonal na antas ng kasiguruhan ng sterility.
- Continuous recording of Temperature (37-55℃).
- Pressure and Humidity (40%-60%RH) pagsubaybay.
- Exposure time (1-3h) and aeration/exhaust time (up to 12h) tracking.
- Compliance with equipment operation manuals.
- External indicators on every package to confirm process exposure.
- Internal indicator cards placed in the “most-difficult-to-sterilize” locations.
- Verification of color change (typically from Brown to Green).
- Performed for every sterilization batch.
- Utilizes Bacillus stearothermophilus spore variants.
- Indicators are cultured immediately post-cycle.
- The gold standard for determining absolute sterilization efficacy.
4. Post-sterilization Treatment
Once the cycle is complete, strict handling protocols must be followed to maintain sterility: