High-Performance Industrie-ETO-Sterilisator & Systems
BOCON ETO Sterilizers are engineered for the rigorous demands of the global medical device industry. Our systems provide a fully automated, highly efficient, Und 100% safe environment for the sterilization of heat and moisture-sensitive medical products.
Aus 1m³ laboratory units to 100m³ large-scale industrial chambers, we deliver more than just equipment—we provide a complete EO Sterilization System designed to meet ISO 11135 Und CE Standards.
Unmatched Technical Advantages
5-Year Warranty
We provide a 5-year warranty for the entire machine, demonstrating absolute confidence in our equipment's stability.
Top-Tier Engineering
Led by Chief Engineer Mr. Sheng (15+ yrs experience), expert in high-rise automation and thermal power control.
Global Leadership
Preferred supplier for Medline, Cardinal, and Medtronic, holding 90%+ share among China's Top 3 manufacturers.
Advanced Control System
- Unrivaled Stability: Unsere Softwarefunktionen 400+ auto-assessment functions and 120+ predictive alarms.
- Full Traceability: Designed to pass the most rigorous audits from Medtronic and Cardinal Health.
- Smart Integration: Standardized MES interfaces for cloud data backup and remote diagnostic capabilities.
Fig 4.1: BCN-GE Series Automatic EO Control System Interface
Superior Craftsmanship
- Argon Shielded Welding: We use high-purity Argon protection welding to ensure structural longevity and vacuum integrity.
- Premium Components: Unlike competitors, we utilize proprietary Ex-sensors and high-performance burners for maximum safety.
- Quality over Price: Our higher price reflects meticulous design and industrial-grade hardware stability.
BOCON Industrial Welding Standard with Argon Protection
Proven Industrial Scale
- Turnkey Capability: We provide full installation, commissioning, and validation (IQ/OQ/PQ) services globally.
- Continuous Choice: Leading manufacturers repeatedly return to BOCON after testing other domestic alternatives.
- R&D Investment: 2-3 million RMB annual investment with 28 international and domestic patents.
Full-scale Industrial ETO Sterilization System Implementation
Technische Daten des BOCON ETO-Sterilisators
Einhaltung globaler Standards: ISO 11135, IN 1422, CE, ISO 13485. Volle Unterstützung für FETT, SA, DQ, IQ, OQ, PQ Validierung.
Technische Unterstützung
Benötigen Sie kundenspezifische Abmessungen oder automatisierte Förderbandladesysteme für Ihr Unternehmen? EO-Sterilisation Linie?
Konsultieren Sie einen Ingenieur →Validated Sterilization Cycle Process
Loading & Setup
Open the loading door, load the products. All of the doors closing. Appropriate parameters setup. Start the cycle.
Heating
Heating up to 52℃.
Keep warm
The system will keep warm for 10 Min.
Pre-Vacuum
Vacuum to -90kPa according to the user parameters.
Leak testing
Duration is 13 Min.
Air to N2 replacement
Replacing the chamber air to N2 gas, 1-2 times.
Humidifying the chamber
The products will be humidified to up to 55%RH.
Humidity dwelling
Duration is 30 Min.
N2 blanket to the chamber
EO pipe washing and preparing the chamber for the EO gas injection.
EO injection to the chamber
The EO gas is injected into the chamber to reach pressure of 25 kPa.
N2 Cover
The process is to wash EO pipe and let all the EO go into the chamber.
EO Dwelling
It will hold on 2 hours and 10 Min.
EO removal
Removing almost all the EO gas out under vacuum.
N2 washing
Injection of N2 gas into the chamber to replace all the EO gas for safety.
Air washing
Washing the products with air, after the air washing process the EO residue becomes very low.
Aeration
This process has the same function with Aeration room, ensuring the EO residue much less.
Cycle Finished
Pressure goes up to -3.0 kPa~3.0 kPa. Open the unloading door and unload products. Print out history data.
Your Trusted Global ETO-Sterilisator Partner
As a leading industrial ETO sterilizer manufacturer, BOCON provides more than just equipment. We deliver a turnkey ETO-Sterilisationssystem backed by 15+ years of engineering excellence. From initial design and IQ/OQ/PQ validation to lifelong technical support, we ensure your medical device sterilization process is safe, compliant, and efficient.
Global Installation
On-site setup and commissioning by expert engineers.
Full Validation
Comprehensive IQ/OQ/PQ documentation and support.
24/7 Unterstützung
Remote diagnostics and rapid spare parts delivery.
ETO-Sterilisator FAQ
Häufig gestellte Fragen zu unseren industriellen Sterilisationslösungen.
- Hohe Durchdringung: Aufgrund der Penetrationseigenschaften von EO-Gas perfekt für die Sterilisation großer Mengen geeignet.
- Endverpackung: Nach dem endgültigen Verpacken sterilisieren, um Sekundärverschmutzung zu vermeiden.
- Materialsicherheit: Bei der Verarbeitung bei niedriger Temperatur bleiben die ursprünglichen Eigenschaften des Produkts erhalten.
- Globales Vertrauen: Lange Geschichte der sicheren und effektiven Anwendung mit ca. 50.000 validierten Medizinprodukten.
- Geringer Wartungsaufwand: Entwickelt für den industriellen Maßstab mit geringen Betriebs- und Wartungskosten.
Unser Industrieller ETO-Sterilisator Lösungen eignen sich für ein breites Produktspektrum. Für konkrete Beratungen, Bitte Kontaktieren Sie uns.
- Endoskope: Bronchoskope, Zystoskope, Ösophagoskope, Mediastinoskope, usw.
- Kunststoff und Gummi: Handschuhe, Spritzen, Infusionsgeräte, Katheter, und Taschen.
- Instrumente: Herzschrittmacher, Künstliche Knochen, Nähte, Knochenplatten, und medizinische elektrische Bohrmaschinen.
- Anästhesiegefäße: Endotrachealtuben, Masken, Sauerstoffzelte, und Zerstäuber.
- Textilien: OP-Kittel, Kappen, Bettwäsche, Gaze, und Bandagen.
Bereit für den Einsatz eines Zuverlässigen
Industrie-ETO-Sterilisator System?
Ganz gleich, ob Sie eine Einheit im Labormaßstab oder eine große Industrieanlage benötigen ETO-Sterilisationssystem, Unsere Ingenieure sind bereit, Ihnen einen maßgeschneiderten technischen Vorschlag zu unterbreiten.







Rezensionen
Es liegen noch keine Bewertungen vor.