(ОК) Operational Qualification For Eto Sterilizer Equipment

(ОК) Operational Qualification For ETO Sterilizer

Operational Qualification (ОК) is a critical phase in the validation of Ethylene Oxide (ЭО) sterilization equipment, ensuring that the chamber and its systems function according to design specifications and ИСО 11135:2014 стандарты.

What is OQ in ETO Sterilization?

OQ is the rigorous process of validating that the sterilization chamber meets all performance requirements specified in its design. It confirms that the system maintains control over key variables like temperature, pressure, and vacuum levels under specified operating procedures.

How is the OQ Procedure Conducted?

The goal is to record that the installed EO sterilizer operates within its predetermined range. Key considerations include:

  • Loading: There is no specific requirement for the chamber to be empty or loaded during OQ; however, parameters should be tested under various conditions.
  • Parameter Stressing: It is recommended to set operating parameters that approach or slightly exceed process limits. This ensures the equipment remains stable even under extreme or challenging operational conditions.
Important: Should we perform PQ before OQ?

No. Operational Qualification (ОК) must be completed and proven before Performance Qualification (ПК). OQ verifies the equipment’s fundamental functionality. If OQ fails, it indicates the equipment is unreliable, making any subsequent PQ data invalid and unverified.

Regulatory Compliance

Adhering to ИСО 11135-1:2014, BOCON provides comprehensive OQ documentation and support to help medical device manufacturers achieve successful system validation. Our protocols ensure that every cycle is repeatable, safe, and fully documented for regulatory audits.